Demand Mode Pacemaker's Ability to Sense ECG
Activity
The two following, related tests, PRP and SRP, determine the demand mode
pacemaker's ability to sense ECG activity immediately following either a
paced event (PRP) or a sensed ECG event (SRP).
1. From the SELECT TEST menu, press F1 REFRACT DEMAND.
2. Select either of the two options:
F1 ATR for the atrial channel.
F2 VENT for the ventricular channel.
4-20
F1
F2
F1
F2
F3
F4
F3
F4
eyr033.eps