Fluke SigmaPace 1000 Operator's Manual page 123

External pacemaker analyzer
Hide thumbs Also See for SigmaPace 1000:
Table of Contents

Advertisement

Non-Capture
Non-Function
3. Press F1 NEXT to advance to the following screen:
4. Press F1 PR to adjust the PR Interval.
5. Press F2 THCH to toggle between atrial and ventricular of the threshold
amplitude.
6. Press F3 THRESH UP or DOWN arrow key to adjust the selected atrial
or ventricular threshold amplitude.
7. Press F4 NSR RATE UP or DOWN arrow key to adjust the NSR rate of
the simulation ECG signal.
8. Conduct pacemaker testing, using the Analyzer to provide the ECG signal.
9. Press ESC to return to previous menu(s).
Transvenous Pacemaker Testing
F1
F2
Setup and Testing
F3
F4
4
eyr048.eps
4-37