Dc Millivolts Measurement Test - Fluke ProcessMeter 789 Calibration Manual

Hide thumbs Also See for ProcessMeter 789:
Table of Contents

Advertisement

789/787B
Calibration Manual

DC Millivolts Measurement Test

1. Put the calibrator in Standby (STBY) mode.
2. Put the UUT rotary switch in the
3. Connect the calibrator to the COM and
4. Apply the values from the calibrator shown in Table 9 and compare the readings on
Range
No Range Switching
No Range Switching
22
Figure 8.
the UUT to the acceptable readings shown.
UUT
Figure 8. DC mV Measurement Test Connections
Calibrator DC Voltage
100 mV
300 mV
U
position.
terminals on the UUT as shown in