Pre-use Inspections

Be sure to perform the following inspections before you use the processor. If you discover any abnormality, stop using the
processor immediately and contact our service personnel.
When using the processor for the very first time, open the lamp housing cover and check the followings:
• Both of two lamps are securely attached
• Lamp socket is not disconnected
• Position of slide plate for lamp mounting is not out of alignment.
Check the following before you begin the inspections:
• The processor is turned off.
• The processor is installed in a stable and level location.
• The scope is connected properly.
• The lamp life indicator is not orange.
• The water bottle is prepared properly and connected correctly.
• The keyboard is connected properly.
3
WARNING
Be sure to supply power to the processor and peripheral devices from the SAT-1300 isolation transformer specified by
PENTAX or another medical isolation transformer.

Power and Lamp Inspection

1
Turn on the monitor and peripheral devices.
2
Press the power switch to turn on the processor and make sure the front panel lights illuminate.
• If pressing the power switch does not turn on the processor, check the fuse.
I "Replacing the Fuses" (P.46)
3
Press c.
4
Make sure the Lamp light illuminates and the distal end of the scope emits light.
• You can have the lamp light come on based on one of the three following configurations:
Start: The lamp lights when the processor is turned on.
Off: The lamp remains off after turning on the process until the lamp switch is pressed.
Lock Lever: The lamp light will come on once the locking lever is moved all the way to the LOCK position.
I [Menu] Key
5
Make sure that the SERVICE lights 1 and 2 are green.
The lamp life for the EPK-p processor is 50 hours. Check the service lamp life indicator on the front panel before you
use the processor. If the service light is orange for either service light 1 or 2 and "Please replace lamp (1 or 2)"
appears on the monitor, replace the affected lamp.
I "Replacing the Lamp" (P.44)
Do not look directly at the light emitted from the distal end of the scope.