4-2 PRE-USE SAFETY CHECKLIST

WARNING - Before every use the following points should be
checked. If any function or device in the video endoscope system
does not perform properly, DO NOT perform the endoscopic
examination. Contact the manufacturer of the device, your
Pentax sales representative or a Pentax service center before
using the equipment for an endoscopic examination.
1. Ensure video processor is placed in stable and level position.
2. Ensure water bottle is properly prepared and connected.
3. Ensure endoscope (and fiberscope video adaptor module if
applicable) is connected properly.
4. Ensure EPK-700 lamp life is less than 200 hours. If lamp life
is 200 hours or more, contact your Pentax sales representative
or a Pentax service center to replace the lamp.
5. Turn ON monitor and other peripheral devices.
6. Turn ON EPK-700 power switch. Ensure switch is lighted
green, the lamp is ignited and the sound of the ventilation
fans can be heard.
CAUTION
- Always turn ON the processor AFTER connecting
an endoscope.
NOTE
Power function switches should not be motivated by contact
with objects like endoscope light guides, when being switched
ON or OFF.
When power is turned OFF, wait 60 seconds before turning
power switch ON. Failure to do so may compromise the
functionality.
Before turning the EPK-700 power ON, ensure the air flow
vents are not obstructed.
7. If lamp fails to ignite, turn EPK-700 OFF, wait 60 seconds
and repeat steps 6 and 7.
NOTE
-If lamp fails to ignite, DO NOT attempt to perform an
endoscopic examination, contact your Pentax service facility.
8. With lamp lit and endoscope connected, check for live
endoscopic image on the monitor.