Cleaning Image Plates Of The Cr Detectors, Plates And Cassettes - AGFA CR HD5 Series User Manual

Cr detectors, plates and cassettes
Table of Contents

Advertisement

50 | AGFA CR Detectors, Plates and Cassettes (CR HD5.x, CR MD4.xR, CR HM5.x, CR MM3.xR) |
Handling the AGFA CR Detectors, Plates and Cassettes

Cleaning Image Plates of the CR Detectors, Plates and Cassettes

The inner lining of the CR detectors is made of fleece. This ensures a high
degree of protection against electrostatic charging and dust collection on the
image plates. Nonetheless, it is recommended to clean the image plates of the
CR HD5.x detectors and CR MD 4.xR plates and cassettes quarterly using the
following procedure.
The CR MM3.xR image plate and the image plate of a CR HM5.x Mammo
detector requires a more frequent cleaning: at least once a week or after every
200 cycles (whatever comes first).
1. Open the shutter of the cassette with the dedicated key.
2322D EN 20180409 1008
WARNING:
Be aware that the image plate of the CR HD5.x and CR HM5.x
detectors is a very expensive part and therefore needs extra
care!
WARNING:
Do not put the image plate with the phosphor side facing down!
WARNING:
To avoid any deformation always store the image plate
horizontally on a flat surface.
WARNING:
Do not use AGFA CR Phosphor Plate Cleaner or any alcohol
containing liquid on the cassette surface, as this can cause
damage to the cassette.
WARNING:
Do not use disinfectants or any other solution on the image
plate.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr md4 r seriesCr hm5 seriesCr mm3 r seriesCr hd5 flfs seriesCr hd5 aec seriesCr md4 r flfs series

Table of Contents