18 | AGFA CR Detectors, Plates and Cassettes (CR HD5.x, CR MD4.xR, CR HM5.x, CR MM3.xR) |
Introduction to the AGFA CR Detectors, Plates and Cassettes

Compatibility

The use of CR HM5.x Mammo detectors is restricted by compatibility criteria
specified in the table below, depending on the availability of an IP gain
calibration file.
There are two valid system configurations. If one component is upgraded or
replaced, all other components must be upgraded or replaced to fit the new
configuration.
Table 1: Compatibility of CR HM5.x Mammo detectors with the NX software
Configuration 1
NX software
CR HM5.x Mammo
detector
Configuration 2
NX software
CR HM5.x Mammo
detector
How to retrieve the version number of the NX software is described in the NX
User Manual.
2322D EN 20180409 1008
Version NX 2.0.8500/3.0.8500 or older
Without IP gain calibration file
Version NX 2.0.8500/3.0.8500 SU1 or newer. For
plates or detectors that are delivered with an IP gain
calibration file, the IP gain calibration files must be
loaded on the NX workstation.
With or without IP gain calibration file