9. When the image plate is completely inside, proceed as follows:
•
First insert the key into the cassette.
•
Then, close the shutter.
10. Remove the key.
Handling the AGFA CR Detectors, Plates and Cassettes | 53
WARNING:
Make sure to slide in the image plate carefully. Do not let the
plate fall down vertically into the cassette! This causes
delaminations and damages of the phosphor.
Note: Make certain you put the correct image plate into the
correct cassette. The label on the image plate must
correspond to the label on the cassette.
Note: After cleaning it is necessary to erase the detector before
use.