Introduction to the AGFA CR Detectors, Plates and Cassettes
Proce-
dure
Table 4: Finding the right gain calibration file
Problem
Solution
Procedure 1. Read the RF tag on the plate or detector. The field "Identifi-
2322D EN 20180409 1008
1. On the NX workstation, click Read and Initialize Cassette in the
Functionality Overview pane of the Main Menu window.
2. Insert a cassette in the ID Tablet.
3. Click Read.
The Read and Initialize Cassette pane shows the details of the in-
serted cassette.
On NX 2.0.8500/3.0.8500 or newer, the field "IP Gain Calibration"
specifies if the plate or detector requires a gain calibration file.
On NX 2.0.8400/3.0.8400 or older, an error is displayed in case the
plate or detector requires a gain calibration file ("The cassette is
not initialized (properly)"). The plate or detector can only be used
after upgrade of the NX software.
It is not clear which IP gain calibration CD belongs to which
plate or detector.
Match the identification values.
cation" contains the plate identification.
2. Check whether the "Identification" on the label of the CD
corresponds with the plate identification.