AGFA CR Detectors, Plates and Cassettes (CR HD5.x, CR MD4.xR, CR HM5.x, CR MM3.xR) |
2. Remove the key.
3. Turn the cassette around, so that the black tube side faces upwards.
4. Fix the shutter with both thumbs and let the image plate slide out carefully
onto the table.
5. Put the image plate onto the black tube side of the empty cassette, as
shown in the figure below.
By putting the plate on the cassette so that the hooks are hanging over the
border of the cassette, a bending of the image plate is avoided.
6. For CR MD4.xR and CR HD5.x, only use AGFA CR Phosphor Plate Cleaner
and a soft lint-free cellulose cloth or Polynit wipes to clean the plate.
For CR MM3.xR and CR HM5.x, only use AGFA CR Phosphor Plate Cleaner
and Polynit wipes to clean the plate.
Handling the AGFA CR Detectors, Plates and Cassettes | 51
Note: You can also use a pen to open the shutter of the
cassette.
2322D EN 20180409 1008