IMMUNITY test
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Power
frequency(50/60 Hz)
magnetic fieldIEC
61000-4-8
Note: UT is the a.c. mains voltage prior to application of the test level.
Declaration – electromagnetic immunity
The Philips Avent breast pump is intended for use in the electromagnetic environment specified
below. The user of the Philips Avent breast pump should assure that it is used in such an environment.
IEC 60601 test level Compliance level
<5 % UT (>95 % dip in
UT) for 0,5 cycle40 %
UT (60 % dip in UT)
for 5 cycles70 % UT
(30 % dip in UT) for 25
cycles<5 % UT (>95 %
dip in UT) for 5 s
3 A/m
<5 % UT (>95 % dip in
UT) for 0,5 cycle70 %
UT (30 % dip in UT) for
25 cycles<5 % UT (>95
% dip in UT) for 5 s
3 A/m
English
Electromagnetic
environment –
guidance
Mains power quality
should be that of a
typical residential or
hospital environment.
If the user of the Philips
Avent breast pump
requires continued
operation during power
mains interruptions,
it is recommended
that the Philips Avent
breast pump be
powered from an
uninterruptible power
supply or a battery.
Interruptible electric
supply can cause the
breast pump to turn
off. This is acceptable
as this will not lead to
unacceptable risks.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical residential or
hospital environment.