Backup/restore of the configuration

The imaging system configuration should be backed up following every service call to
ensure quick system restore.
Advantage:
-
current customer organ programs
-
current configuration files.
Included in this backup are images in extended storage, organ programs, configuration
data of the imaging system.
The local network data from the operating system is not stored (e.g. IP address, subnet
mask, gateway IP address).
From the service tool of the imaging system, all backup data is stored in the d:/_Siemens_
folder. Save data to CD in Windows XP by selecting "Write these files to CD" in Internet
Explorer.

Backup of the configuration

Preparation:
Prior to every backup, perform a check to determine whether the images in extended stor-
age have been deleted or made anonymous.
Procedure
Switch off the MOBILETT XP Digital.
Open the top cover of the MOBILETT XP Digital, lay it sideways on a table or the like
and access the PC interior. Do not disconnect any cables.
Open the PC so that it is possible to use the PC keyboard.
Switch on the Mobilett XP Digital.
From the user interface of the MOBILETT XP Digital:
Select "SYSTEM > CONFIGURATION".
Select "ADMINISTRATOR SETUP > SERVICE TOOL".
Click on the password box
Select "SERVICE TOOL > Copy files to _Siemens_ " in the menu.
Start via "START".
Close the window via "OK".
Burn data to CD
MOBILETT XP Digital
- Enter the administrator password and confirm with "OK" (refer to the password
list).
Enter the Siemens service password and confirm with "OK" (refer to the pass-
word list).
All required files are now copied to the D:\_Siemens_ directory.
SPR8-230.840.30.03.02
12.05
Backup and restore
Page 60 of 68
CS SD 24
0
0
Siemens AG
Medical Solutions