52
Tab. 10
Customer Attention Table
Error Code
-506
-508
-511
MOBILETT XP Digital
Error messages of the imaging system
Error Message
DICOM Response Warning:
Although transfer finished prop-
erly, receiving device reported a
warning in DICOM response.
Check receiving device.
DICOM Printer Status Warning:
Although transfer finished nor-
mally, receiving device reported
a WARNING in DICOM printer
status. Check receiving device.
External Storage Attention:
Remaining space in external
storage is less than **% .
SPR8-230.840.30.03.02
12.05
Remedy
•
Touch OK and check the settings of
the output destination.
•
Touch OK and replace the external
disk before the capacity becomes
insufficient. ** is the value you have
set.
Page 52 of 68
CS SD 24
Siemens AG
Medical Solutions