16

Fatal error

Tab. 1
Service Fatal Error Table
Code
-100
-102
-103
MOBILETT XP Digital
Error messages of the imaging system
Reply
Message
keys
OK,
#1:Fatal error
#2:Sensor unit: Self-check error
#3:Detected an error.#4:ER(n) (xxxx) xxxx: Error code
[Cause]
An undefined error has been found in the sensor A/D board.
[Operation after the reply key is pressed]
Nothing
-14,
-14
-15,
Error in starting the task in DLL (failed to begin thread.)
-95,
-98
-15
Error in DLL resource (driver does not start.)
-95
Number of sensor units is incorrect (no sensor unit is con-
nected.)
-98Capture device cannot be found.
Shut
#1:Fatal error
down
#2:READ timeout error
#3:Failed to transmit the data.
[Cause]
Image transferring from Ethernet has failed.
[Operation after the reply key is pressed]
For the reply cannot be done, the system must be shutdown.