Error messages of the imaging system
Alert
Tab. 3
Service Alert Table
Code
-6
-8
Siemens AG
Medical Solutions
Reply
Message
keys
OK
#1:Alert
#2:System info error
#3:A/D board info is updated.
[Cause]
The A/D board information is different from the former informa-
tion. This problem occurs when the version of the A/D board has
been updated or when the connected units have been changed.
If this error occurs when no such changes have been made, it is
attributed to another problem. In this case, a service person
needs to be called in.
[Operation after the reply key is touched] The A/D board infor-
mation is renewed.
OK
#1:Alert
#2:System self-check error
#3:Call a service person and let him/her know the error code.
[Cause]
An error has been found by a self-check.
[Operation after the reply key is pressed]
Although imaging is possible, image may be degraded. There-
fore, a service person needs to be called in.
SPR8-230.840.30.03.02
12.05
MOBILETT XP Digital
Page 19 of 68
CS SD 24
19
0