Error messages of the imaging system
Code
-513
-514
Siemens AG
Medical Solutions
Reply
Message
keys
OK
#1:Error
#2:Media control error
#3:Operation has been canceled because the external storage
or directory is busy.
[Cause]
The external medium is busy deleting or writing data. This error
occurs for example when you try to eject the disk while the
external medium is busy.
[Operation after the reply key is pressed]
After confirmation, press OK.
OK
#1:Error
#2:External storage error:
#3:The exposed images cannot be stored because the size of
the folder partition is smaller than the exposed images. Use a
new disk, erase the folder contents or enlarge the folder size
and retry.
[Cause]
An image cannot be stored in a folder on the external storage
medium when the size of the folder is smaller than the image.
Enlarge the folder size in the system setting.
[Operation after the reply key is pressed]
Selecting RETRY, stores the image again, while selecting CAN-
CEL cancels storage of the image to the external storage
medium. When CANCEL is selected, a confirmation message
appears before the image is deleted.