Siemens MOBILETT XP Digital Troubleshooting Manual page 45

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

Error messages of the imaging system
Code
-514
-515
Siemens AG
Medical Solutions
Reply
Message
keys
RETRY
#1:Warning
CANCEL
#2:External storage error:
#3:The exposed images cannot be stored because the size of
the folder partition is smaller than the exposed images. Use a
new disk, erase the folder contents or enlarge the folder size
and retry. The size of the folder partition: xxxxMB) (The size of
the exposed images: xxxxMB)
[Cause]
Images cannot be stored in the folder when the folder size of the
external storage medium is smaller than a study.
[Operation after the reply key is pressed]
When RETRY is selected, the image will be stored. When CAN-
CEL is selected, the storing of the image will be cancelled after a
confirmation message is displayed. Re-transfer is possible by
transferring the image from the study list.
RETRY
#1:Warning
CANCEL
#2:Internal storage error:
#3:Selected directory for storage is read-only.
[Cause]
The external storage medium is read-only.Use a writable exter-
nal storage medium.
[Operation after the reply key is pressed]
When RETRY is selected, the image will be stored on the inter-
nal disk.When CANCEL is selected, the storing of the image will
be cancelled after a confirmation message is displayed.
Re-transfer is possible by sending the image from the study list.
SPR8-230.840.30.03.02
12.05
MOBILETT XP Digital
Page 45 of 68
CS SD 24
45