14
3-
General
2Error messages of the imaging system
Service messages and messages from the operator's manual (customer messages) are
included.
The following table shows the alert code window.
Fig. 7:
Area
#
1
2
3
MOBILETT XP Digital
Error messages of the imaging system
Alert window
Severity
Auto-
matic
display
Fatal
Yes
error
Alert
Yes
Error
Yes
SPR8-230.840.30.03.02
12.05
Blinking
Meaning
of "ERR"
Yes
The whole screen blinks red and the alert
window appears automatically. The system
cannot be recovered. A service person must
be called.
Yes
A red ERR key blinks and the alert window
appears automatically. The exposure can-
not be continued without a user operation.
No
A red ERR key is displayed and the alert
window appears automatically. The ERR key
does not blink.The chances of the user being
able to recover the system is better than for
an alert.