Error messages of the imaging system
Reply
Message
keys
OK,
#1:Error
0xeff3,
#2:Sensor unit:Control error
0xeff4
#3:Detected control error. Call a service person.
[Cause]
An error in the control line has been detected in the sensor A/D
board.
[Operation after the reply key is pressed]
When the OK key is pressed, the sensor will be disconnected.
(Access will not be allowed, as the system is in an "abnormal"
state.) A service person needs to be called in.
0xeff3, 0xeff4
"Sensor unit - power box(n)(eff3 or eff4)" 0xeff3: There might be
problems the cable length setting (sensor power box and X-ray
board cable). 0xeff4: The sensor power cable may be bro-
ken.
n: Sensor index (1 to 4)
OK, can-
#1:Error
cel,
#2: Sensor unit: PLD error
0xeff8,
#3: Detected PLD error. Call a service person.
0xeff7
[Cause]
A PLD error has been found in the sensor A/D board.
[Operation after the reply key is pressed]
When the OK key is pressed, the sensor will be disconnected.
(Access will not be allowed, since the system is in an "abnormal"
state.) If this error persists even after the system is rebooted, a
service person needs to be called in.
0xeff8, 0xeff
"PLD err (n)( eff8)""PLD Config err (n)( eff7)"
n: Sensor index (1 to 4)