MOBILETT XP Digital
Error messages of the imaging system
Error Message
Detector Unit: Signal Line Error:
Problem has occurred between
detector unit and EO box. Call a
service person.
Capture Board: Data Transmit
Error: Failed to transmit the data.
Call a service person.
Capture Board Error: Detected
capture board error.
Capture Board Control Error:
Detected capture board control
error. Call a service person.
Capture Board Error: Detected
capture board error. Call a ser-
vice person.
Capture Board Access Error:
Detected Communication Error.
Capture Board: Sequence Error:
Failed to synchronize with the
detector unit. Call a service per-
son.
Capture Board: Sequence Error:
X-ray monitor in detector unit
failed to detect start of radiation.
Capture Board: Data Transmit
Error: Failed to transmit data.
Call a service person.
Capture Board: Control Line
Error: Control line error has
occurred in capture board. Call a
service person.
Acquisition Parameter Error:
Detected illegal parameter error.
LUT Write Error: Quit the exam
and perform calibration.
Calibration Table Error: Perform
calibration.
SPR8-230.840.30.03.02
12.05
Remedy
Touch OK and shut down the sys-
tem.
If this error occurs, contact a Sie-
mens representative or distributor.
Touch OK to close the window.
As image quality cannot be guaran-
teed, take another exposure.
Touch OK and shut down the sys-
tem.
If these errors occur, contact a Sie-
mens representative or distributor.
Touch OK and calibrate the unit.
Page 36 of 68
CS SD 24
Siemens AG
Medical Solutions