Siemens MOBILETT XP Digital Troubleshooting Manual page 17

Hide thumbs Also See for MOBILETT XP Digital:
Table of Contents

Advertisement

Error messages of the imaging system
Code
-105
-522
Siemens AG
Medical Solutions
Reply
Message
keys
OK/ -97
#1:Fatal error
#2:CXDCAP.INI Open error
#3:Install CXDCAP.INI file.
[Cause]
The cxdcap.ini file has not been installed.
[Operation after the reply key is pressed]
The system enters a fatal condition. All the sensor units will
therefore be disconnected. A service person needs to be called
in. However, you can transfer the image and use the study list.
-97
Cxdcap.ini file not found.
OK
#1:Fatal error
#2:File Open Error:filename
#3:File is not found.
[Cause]
For some reason, XXXX.BIN and XXXX.DCM files do not exist.
Or, they cannot be read.
[Operation after the reply key is pressed]
Shutdown starts. Although this problem can be solved using the
service tool, users must not use the service tool on their own.
Take necessary actions in accordance with the service person"s
instruction.
SPR8-230.840.30.03.02
12.05
MOBILETT XP Digital
Page 17 of 68
CS SD 24
17