98 | DX-D 40C, DX-D 40G | Problem solving

DR detector not ready for exposure

Details
Cause
Cause
Cause
Brief Solution
Related Links
Registering the DR Detector on another NX Workstation
Using the Windows wifi settings to switch between the wireless DR detector and
the wireless hospital network
0290B EN 20190401 1627
The DR detector is turned on. The DR Detector Status
icon on the DR Detector Switch is not green.
(only if the DR Detector is shared between multiple NX
workstations)
The DR Detector is not registered on the NX Worksta-
tion.
(only on DX-D 45C, DX-D 45G)
The S-button has been pushed accidentally.
(only if the DR Detector communicates via the internal
wireless adapter of the workstation)
The NX Workstation may not be connected to the DR
Detector via the wireless network.
1. (only if the DR Detector is shared between multiple
NX workstations)
Register the DR Detector on the NX Workstation.
2. (on DX-D 45C, DX-D 45G)
Check the indicator next to the S-button. The indi-
cator must be green or orange when the DR Detec-
tor communicates via a System Control Unit or a
wireless access point that is connected to the work-
station. The indicator must be blue when the DR
Detector communicates via the internal wireless
adapter of the workstation.
If the indicator does not match the way it communi-
cates to the workstation, press and hold the S-but-
ton for 5 seconds.
The indicator color switches to the correct mode.
3. (if the DR Detector communicates via the internal
wireless adapter of the workstation)
Switch the NX Workstation to the wireless network
of the DR Detector.
on page 94
on page 92