104 | DX-D 40C, DX-D 40G | Technical Data

DX-D 40C, DX-D 40G

Commercial name
Electrical connection DR Detector
Rated power supply
(powered by battery pack)
Power consumption
Wireless connection
Wireless signal range (in an open
space)
Environmental conditions (during normal operation)
Room temperature
Humidity (non condensing)
Atmospheric pressure
Environmental conditions (during storage and transport)
Temperature (ambient)
Humidity (non condensing)
Atmospheric pressure
Warming-up time
30 minutes
Dimensions
Dimensions
width x length x height
Weight (incl. battery)
Maximum load
Maximum total load
0290B EN 20190401 1627
DX-D 40C, DX-D 40G
DC +24 V, Max. 0.5 A
max. 12 W
IEEE 802.11a/b/g/n (2.4 GHz/5 GHz)
maximum 8 m
between +10 °C and +35 °C
between 30% and 85% RH
(non-condensing)
between 700 hPa and 1060 hPa
between -15 °C and +55 °C
between 10% and 90%
(non-condensing)
between 500 and 1060 hPa
approx. 384 x 460 x 15.5 mm
< 3.4 kg
100 kg on an area of 40 mm in diameter
150 kg over the whole detector surface