72 | DX-D 40C, DX-D 40G | Getting started

Starting the DR Detector

To start the DR Detector:
1. If the DR Detector is connected to the System Control Unit via the DR
Detector cable, check if the power cable of the System Control Unit is
connected to the mains power.
The battery is not required to operate the DR Detector. Skip to step 4.
2. Fully charge the battery.
Charge the battery on the day of examination or on the previous day.
3. Attach the battery.
Align the claw on the battery pack and the groove on the battery bay (1).
Insert the battery pack fully (2). Push down the battery pack (3). Slide the
lock lever toward (lock) side (4) and lock it.
1
Figure 13: Attach the battery
0290B EN 20190401 1627
CAUTION:
Do not use the battery pack as a power source for equipment
other than DX-D 40C or DX-D 40G detectors. Be sure to use only
the dedicated battery pack for the DX-D 40C or DX-D 40G
detector.
Note: The battery slowly discharges even if not in use. The
battery pack may have expired if it discharges immediately
after being fully charged. You can purchase an optional battery
pack to replace an exhausted one.
Note: Make sure that the lock lever is placed to the (unlock)
side.
1
2
Note: Make sure that the battery is securely attached.
3
4
4