36 | DX-D 40C, DX-D 40G | Introduction to the DR Detector

Wired communication

The use of accessories and cables other than those specified or sold by the
manufacturer as replacement parts, may result in increased radiation
emissions or decreased stability of the equipment.
Accessory equipment connected to the analog and digital interfaces must be
certified according to the respective IEC standards. All combinations of
equipment must be in compliance with IEC 60601-1-1 system requirements.
Any person who connects additional equipment to the signal input or signal
output ports, configures a medical system and is therefore responsible for
ensuring that the system complies with the requirements of the system
standard IEC 60601-1.