Charging a battery

To charge a battery using the battery charger:
1. Connect the power supply to the mains power and to the power socket of
the battery charger.
2. Insert the battery in an empty slot of the battery charger.
The battery charger automatically detects the battery and starts charging
the battery.
The battery status can be read from the indicator lights.
The charge level of the battery is monitored and it is kept at maximum
level until the battery is removed from the battery charger.
3. Remove the charged battery from the battery charger.
Related Links
Safety Directions
DR Detector Battery Charger

Charging the battery using the DR detector cable

Connect the DR Detector cable to charge the battery that is attached to the DR
Detector. The battery status can be read from the DR Detector Switch on the
NX Workstation.
While charging, the DR Detector can still be used. If the DR Detector is used
with the DR Detector cable connected, make sure that a battery is always
attached.
Related Links
DR Detector Switch on the NX Workstation