92 | DX-D 40C, DX-D 40G | Advanced Operating
Registering the DR Detector on another NX
Workstation
The DR Detector can be used for examinations on different NX workstations.
The DR Detector is set up to communicate to a specific NX workstation. The
procedure of registering the DR Detector to another NX workstation switches
the availability of the DR Detector between NX workstations.
In a configuration with multiple NX workstations sharing a DR Detector, each
NX workstation is connected to a System Control Unit. One or more of the
System Control Units is equipped with a DR Detector Cable.
In a configuration with mobile X-ray units sharing a DR Detector, a dedicated
PC is connected to a System Control Unit that is equipped with a DR Detector
Cable.
To register the DR Detector and make a connection to another X-ray room:
1. Connect the DR Detector to any of the NX workstations with the DR
Detector Cable.
A dialog is displayed on the NX workstation listing the configured X-ray
rooms.
It may take up to 30 seconds for the dialog to pop up.
2. Select the X-ray room where the DR Detector will be used.
A dialog is displayed on the NX workstation to confirm the registration.
0290B EN 20190401 1627
Note: On a PC without NX software, the VDI Panel Sharing Tool
may not be set up to start automatically. To start it manually, in
the Start menu, select All programs > Agfa > Start VDI Panel
Sharing Tool.