Description of Device Functions
Proline t-mass 65
MODBUS RS485
Thermal Mass Flow Measuring System
BA00116D/06/EN/13.10
71123867
Valid as of version
V 3.06.XX (Device software)

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline t-mass 65 MODBUS RS485

  • Page 1 Description of Device Functions Proline t-mass 65 MODBUS RS485 Thermal Mass Flow Measuring System BA00116D/06/EN/13.10 71123867 Valid as of version V 3.06.XX (Device software)
  • Page 3: Table Of Contents

    Proline t-mass 65 MODBUS RS485 Table of contents Table of contents The function matrix: layout and use . . . 4 Illustration of the function matrix..5 MEASURING VALUES ... . . 7 SYSTEM UNITS .
  • Page 4: The Function Matrix: Layout And Use

    The function matrix: layout and use Proline t-mass 65 MODBUS RS485 The function matrix: layout and use The function matrix is a two-level construction: the groups form one level and the groups' functions the other. The groups are the highest-level grouping of the control options for the measuring device.
  • Page 5: Illustration Of The Function Matrix

    Proline t-mass 65 MODBUS RS485 Illustration of the function matrix Illustration of the function matrix Groups/function groups Functions CORRECTED VOLUME  ä 7 MEASURING VALUES MASS FLOW HEAT FLOW TEMPERATURE FLOW  UNIT CORRECTED UNIT CALORIFIC VALUE UNIT CALORIFIC VALUE ...
  • Page 6 Illustration of the function matrix Proline t-mass 65 MODBUS RS485 Groups/function groups Functions  ä 39 SELECT GROUP NUMBER OF GASES GAS TYPE 1 MOLE % GAS 1 DESCRIPTION CORRECTION FACTOR REFERENCE DENSITY GAS TYPE 2 to 8  MOLE % GAS 2 to 8...
  • Page 7: Measuring Values

    Proline t-mass 65 MODBUS RS485 MEASURING VALUES MEASURING VALUES Function description, MEASURING VALUES group Note! The engineering unit of the measured variable displayed here is configured in the SYSTEM UNITS group ( ä 8) MASS FLOW Description The currently measured mass flow appears on the display.
  • Page 8: System Units

    SYSTEM UNITS Proline t-mass 65 MODBUS RS485 SYSTEM UNITS Function description, SYSTEM UNITS group UNIT MASS FLOW Description For selecting the unit required and displayed for the mass flow. The unit you select here is also valid for: MODBUS register: 2101 •...
  • Page 9 Proline t-mass 65 MODBUS RS485 SYSTEM UNITS Function description, SYSTEM UNITS group UNIT CALORIFIC VALUE Description CORRECTED VOLUME For selecting the unit required and displayed for the calorific value based on the corrected volume. MODBUS register: 2156 Options (SI units):...
  • Page 10 SYSTEM UNITS Proline t-mass 65 MODBUS RS485 Function description, SYSTEM UNITS group UNIT DENSITY Description For selecting the unit required and displayed for the calculated gas density at process conditions. The unit you select here is also valid for: MODBUS register: 2107 •...
  • Page 11: Quick Setup

    Proline t-mass 65 MODBUS RS485 QUICK SETUP QUICK SETUP Function description, QUICK SETUP group QUICK SETUP Description COMMISSIONING Starts the Quick Setup menu for commissioning. For a flowchart of the QUICK SETUP COMMISSIONING:  ä 13 Options 0 = NO...
  • Page 12 QUICK SETUP Proline t-mass 65 MODBUS RS485 Function description, QUICK SETUP group QUICK SETUP HEAT FLOW Description Use this function to start the application-specific setup for heat flow. For a flowchart of the QUICK SETUP HEAT FLOW:  ä 20...
  • Page 13 Proline t-mass 65 MODBUS RS485 QUICK SETUP Quick Setup "Commissioning" XXX.XXX.XX Quick Setup Commission Language HOME-POSITION Pre-setting Selection pre-settings Delivery settings Actual settings Selection system units Pressure Temperature Mass flow Corr. Vol. Flow Heat Flow Density Quit Unit Unit Unit...
  • Page 14 QUICK SETUP Proline t-mass 65 MODBUS RS485 SYSTEM UNITS Select required system unit function and carry out parameterization or select QUIT to return to the QUICK SETUP function if no further programming is required. n Only units not yet configured in the current setup are available for selection in each cycle.
  • Page 15 Proline t-mass 65 MODBUS RS485 QUICK SETUP Quick Setup "Communication" To establish serial data transfer, various arrangements between the MODBUS master and MODBUS slave are required which have to be taken into consideration when configuring various functions. These functions can be configured quickly and easily by means of the Communication Quick Setup.
  • Page 16 QUICK SETUP Proline t-mass 65 MODBUS RS485 Quick Setup "Sensor" It is essential that the insertion sensor is setup according to the actual pipe or duct and then installed at the calculated insertion depth. This Quick Setup guides you systematically through the procedure to setup the sensor.
  • Page 17 – selected from a list of standard gases or – setup for other suitable types of gases, such as Ozone, using manual correction factors and the option called SPECIAL GAS. This requires application evaluation at the factory - consult your Endress+Hauser representative prior to using this function. Endress+Hauser...
  • Page 18 QUICK SETUP Proline t-mass 65 MODBUS RS485 Setting or viewing the active gas group Go to the function SELECT GROUP ( ä 40) and simply select 1 or 2 and then exit using ESC (Q keys simultaneously). No save function is necessary.
  • Page 19 Proline t-mass 65 MODBUS RS485 QUICK SETUP "Pressure" Quick Setup menu Use this Quick Setup to program the individual process pressure for each gas group. If only one gas group is being used, then only the function PROCESS PRESSURE 1 needs to be programmed, PROCESS PRESSURE 2 can remain with default settings.
  • Page 20 QUICK SETUP Proline t-mass 65 MODBUS RS485 "Heat Flow" Quick Setup menu The device can calculate and output the heat of combustion of common fuel gases such as methane, natural gas, propane, butane, ethane and hydrogen. Use this Quick Setup menu to program the method used to calculate the heating value or calorific value (CV).
  • Page 21 Proline t-mass 65 MODBUS RS485 QUICK SETUP Manual This function allows entry of a user-specific heating value if the required value is different to the following table. Formula Net/lower heating value Gross/upper heating value [Mj/kg] MBtu/lb [Mj/kg] MBtu/lb Hydrogen 119.91 51.56...
  • Page 22 QUICK SETUP Proline t-mass 65 MODBUS RS485 Data backup/transmission Using the T-DAT SAVE/LOAD function, you can transfer data (device parameters and settings) between the T-DAT (exchangeable memory) and the EEPROM (device storage unit). This is required in the following instances: •...
  • Page 23: Operation

    60 seconds following a return to the HOME position. • Programming can also be disabled by entering any number (other than the private code). • Your Endress+Hauser representative can be of assistance if you mislay your private code. Note! This function is only relevant for the local display and does not affect data transmission via the MODBUS communication.
  • Page 24 OPERATION Proline t-mass 65 MODBUS RS485 Function description, OPERATION group STATUS ACCESS Description The access status for this function matrix appears on the display. MODBUS register: 2512 Display Data type: Integer 0 = LOCKED (parameters cannot be modified) Access: read/write...
  • Page 25: User Interface

    Proline t-mass 65 MODBUS RS485 USER INTERFACE USER INTERFACE Function description, USER INTERFACE group ASSIGN LINE 1 Description For assigning a display value to the main line (top line of the local display). This value is displayed during normal operation.
  • Page 26 USER INTERFACE Proline t-mass 65 MODBUS RS485 Function description, USER INTERFACE group 100% VALUE LINE 2 Prerequisite This function is only available if one of the following options was selected in the ASSIGN LINE 2 function ( ä 25): MODBUS register: 2564 •...
  • Page 27 Proline t-mass 65 MODBUS RS485 USER INTERFACE Function description, USER INTERFACE group BACKLIGHT Description For adjusting the background lighting to suit local operating conditions. Entering the value "0" means that the backlight is "switched off". The display MODBUS register: 2566 then no longer emits any light, i.e.
  • Page 28: Totalizer 1/2

    TOTALIZER 1/2 Proline t-mass 65 MODBUS RS485 TOTALIZER 1/2 Function description, TOTALIZER 1/2 group ASSIGN TOTALIZER Description Use this function to assign a measured variable to the totalizer. MODBUS register: • The totalizer is reset to "0" as soon as the option selected is changed.
  • Page 29 Proline t-mass 65 MODBUS RS485 TOTALIZER 1/2 Function description, TOTALIZER 1/2 group RESET TOTALIZER Description Resets the sum and overflow in the totalizer selected. MODBUS register: Options TOTALIZER 1 2608 0 = NO TOTALIZER 2 2808 1 = YES Data type:...
  • Page 30: Handling Totalizer

    HANDLING TOTALIZER Proline t-mass 65 MODBUS RS485 HANDLING TOTALIZER Function description, HANDLING TOTALIZER group RESET ALL TOTALIZERS Description Resets the sums and overflows of both totalizers to "zero" (=RESET). MODBUS register: 2609 If the device is equipped with a status input and if it is appropriately configured,...
  • Page 31: Status Input

    Proline t-mass 65 MODBUS RS485 STATUS INPUT STATUS INPUT Function description, STATUS INPUT group ASSIGN STATUS INPUT Description Use this function to assign a switching function to the status input. MODBUS register: 4301 • Positive zero return is active as long as the level is present at the status input...
  • Page 32 STATUS INPUT Proline t-mass 65 MODBUS RS485 Function description, STATUS INPUT group SIMULATION STATUS INPUT Description Use this function to simulate the status input, i.e. to trigger the function assigned to the status input (see ASSIGN STATUS INPUT function ( ä 31).
  • Page 33: Communication

    Proline t-mass 65 MODBUS RS485 COMMUNICATION COMMUNICATION Function description, COMMUNICATION group TAG NAME Description Use this function to enter a tag name for the device. You can edit and read this tag name at the local display or via the MODBUS communication..
  • Page 34 COMMUNICATION Proline t-mass 65 MODBUS RS485 Function description, COMMUNICATION group BYTE ORDER Prerequisite The transmission sequence must suit the MODBUS master. More information can be found in Operating Instructions BA115D/06 under MODBUS register: 4915 the keyword "Byte transmission sequence". Data type:...
  • Page 35: Process Parameter

    PROCESS PARAMETER PROCESS PARAMETER Function description, PROCESS PARAMETER group PROCESS PRESSURE 1 Prerequisite This function is not available if the IN-SITU CALIBRATION function has been enabled. Refer to your Endress+Hauser service representative for more MODBUS register: 5268 information. Data type: Float...
  • Page 36 REFERENCE DENSITY Prerequisite This function is not available if the IN-SITU CALIBRATION function has been enabled. MODBUS register: 5130 Contact your Endress+Hauser service organization for more information. Data type: Float Access: read Description Use this function to display the calculated reference density (for corrected volume flow measurement).
  • Page 37 Proline t-mass 65 MODBUS RS485 PROCESS PARAMETER Function description, PROCESS PARAMETER group ON-VALUE LOW FLOW CUT Prerequisite This function is not available if OFF was selected in the ASSIGN LOW FLOW CUT OFF function ( ä 36). MODBUS register: 5138...
  • Page 38 PROCESS PARAMETER Proline t-mass 65 MODBUS RS485 Function description, PROCESS PARAMETER group ZERO POINT ADJUST Description Use this function to start automatic zero point adjustment. " MODBUS register: 5121 Caution! Data type: Integer Prior to performing zero point adjustment, observe the exact procedure for zero...
  • Page 39: Gas

    Function description, GAS group Prerequisite This function is not available if the IN-SITU CALIBRATION function is enabled. This function is available again if the in-situ calibration is reset to the factory setting. Contact your Endress+Hauser service organization for more information. Description Use this function to view or change the gas configuration •...
  • Page 40 Proline t-mass 65 MODBUS RS485 Function description, GAS group SELECT GROUP Description Use this function to: • select a gas group for editing MODBUS Register 9798 • set the active gas group manually Datentyp: Integer Zugriff read/write Setting the active gas group: •...
  • Page 41 Proline t-mass 65 MODBUS RS485 Function description, GAS group DESCRIPTION Prerequisite This function is only available if the option SPECIAL GAS is selected in the function GAS TYPE 1 ( ä 40). MODBUS register: 9957 Data type: String (16) Description...
  • Page 42 Proline t-mass 65 MODBUS RS485 Function description, GAS group GAS TYPE 2 to 8 Prerequisite The number of functions available here is dependent upon the setting in the function NUMBER OF GASES ( ä 40). MODBUS register: GAS TYPE 2...
  • Page 43 Proline t-mass 65 MODBUS RS485 Function description, GAS group SAVE CHANGES Description Use this function to control the way entries are saved in the gas group and utilized for flow measurement. MODBUS register: 9743 Data type: Integer Options Access: read/write...
  • Page 44: Heat Flow

    HEAT FLOW Proline t-mass 65 MODBUS RS485 HEAT FLOW Function description, HEAT FLOW group CALORIFIC VALUE TYPE Description Use this function to select the measured variable on which the combustion value is based. MODBUS register: 9977 Data type: Integer Options...
  • Page 45 Proline t-mass 65 MODBUS RS485 HEAT FLOW Function description, HEAT FLOW group REFERENCE COMBUSTION Prerequisite TEMPERATURE This function is not available if the option MANUAL is selected in MODE 1 or MODE 2 ( ä 44). MODBUS register: 9982 Description...
  • Page 46: System Parameter

    SYSTEM PARAMETER Proline t-mass 65 MODBUS RS485 SYSTEM PARAMETER Function description, SYSTEM PARAMETER group POSITIVE ZERO RETURN Description Use this function to interrupt evaluation of measured variables. For example, the output signal should be set to zero flow during operations such as pipe cleaning...
  • Page 47: Sensor Data

    Proline t-mass 65 MODBUS RS485 SENSOR DATA SENSOR DATA Function description, SENSOR DATA group This group of functions contains the essential data relating to the sensor geometry and calibration. Flange version (t-mass 65F): The sensor data cannot be changed and is read only.
  • Page 48 SENSOR DATA Proline t-mass 65 MODBUS RS485 Function description, SENSOR DATA group NOMINAL DIAMETER Prerequisite This function is only available for insertion sensors (t-mass 65I). This function is not available if OTHER was selected in the PIPE STANDARD MODBUS register: 9632 function (...
  • Page 49 Proline t-mass 65 MODBUS RS485 SENSOR DATA Function description, SENSOR DATA group INTERNAL DIAMETER Prerequisite This function is only available for insertion sensors (t-mass 65I). This function is only available if CIRCULAR is selected in the function PIPE MODBUS register: 9604 TYPE (...
  • Page 50 SENSOR DATA Proline t-mass 65 MODBUS RS485 Function description, SENSOR DATA group MOUNTING SET LENGTH Prerequisite This function is only available for insertion sensors (t-mass 65I). MODBUS register: 9635 Description Data type: Float Use this function to enter a value for the length of the mounting set (including...
  • Page 51: Supervision

    Proline t-mass 65 MODBUS RS485 SUPERVISION SUPERVISION Function description, SUPERVISION group ACTUAL SYSTEM CONDITION Description The current system status appears on the display. MODBUS register: 6859 Display Data type: Integer SYSTEM OK or MODBUS register: read The fault/notice message with the highest priority.
  • Page 52: Simulation System

    SIMULATION SYSTEM Proline t-mass 65 MODBUS RS485 SIMULATION SYSTEM Function description, SIMULATION SYSTEM group SIMULATION FAILSAFE Description MODE Use this function to set all inputs, outputs and the totalizer to their error response modes, in order to check whether they respond correctly. During this time, the message #691 "SIMULATION FAILSAFE"...
  • Page 53: Sensor Version

    Proline t-mass 65 MODBUS RS485 SENSOR VERSION SENSOR VERSION Function description, SENSOR VERSION group SENSOR TYPE Description Use this function to view the sensor type. MODBUS register: 7012 Display Data type: Integer 0 = FLOWCELL (t-mass 65F flange sensor) Access:...
  • Page 54: Amplifier Version

    AMPLIFIER VERSION Proline t-mass 65 MODBUS RS485 AMPLIFIER VERSION Function description, AMPLIFIER VERSION group DEVICE SOFTWARE Description Use this function to display the current device software version. MODBUS register: 7277 Data type: String (16) Access: read HARDWARE REVISION Description NUMBER AMPLIFIER Use this function to view the hardware revision number of the amplifier board.
  • Page 55: Factory Settings

    Proline t-mass 65 MODBUS RS485 Factory settings Factory settings Language  ä 23 21.1 Country Language Country Language Australia English Luxembourg French Austria German Malaysia English Belgium English The Netherlands Dutch Czechia Czech Norway Norwegian Denmark English Poland Polish England...
  • Page 56 Factory settings Proline t-mass 65 MODBUS RS485 Nominal diameter Low flow cut off Full scale value Pulse value [mm] [kg/h] [kg/h] [kg/p] 280.00 28000 100.0 500.00 50000 100.0 800.00 80000 100.0 1150.00 115000 100.0 1590.00 159000 100.0 1000 3200.00 320000 100.0...
  • Page 57 Proline t-mass 65 MODBUS RS485 Factory settings 21.3 US units (only for USA and Canada) 21.3.1 Low flow cut off, full scale value, pulse value t-mass F sensor With air at ambient conditions; (without a flow conditioner) Nominal diameter Low flow cut off...
  • Page 58 Factory settings Proline t-mass 65 MODBUS RS485 Unit totalizer 1 and 2  ä 28 21.3.3 Unit Unit Mass flow Corrected volume flow Sm³ Heat flow kBtu 21.3.4 Other Units Unit  ä 45 Ref. combustion temp. °F  ä 47...
  • Page 59: Index

    Proline t-mass 65 MODBUS RS485 Index Index Numerics GAS TYPE 2 to 8 (fct.) ......42 GAS, function group .
  • Page 60 Proline t-mass 65 MODBUS RS485 Index PROCESS PARAMETER, function group ....35 TEMPERATURE (fct.) ......7 PROCESS PRESSURE (fct.) .
  • Page 61 Erklärung zur Kontamination und Reinigung Please reference the Return Authorization Number (RA#), obtained from Endress+Hauser, on all paperwork and mark the RA# clearly on the outside of the box. If this procedure is not followed, it may result in the refusal of the package at our facility.
  • Page 62 www.endress.com/worldwide BA00116D/06/EN/13.10 71123867 FM+SGML6.0 ProMoDo...

Table of Contents