The LEVel:SRESistance or VOLTage:SRESistance command sets the source
output impedance.
Command
Syntax
Parameter
Example
Query Syntax
Query Response
Example
1.800.561.8187
LEVel:CURRent {<value>,MIN,MAX}
or CURRent:LEVel {<value>,MIN,MAX}
Where,
<value> Is the numeric data (NR1, NR2 or NR3).
MIN Sets to the minimum value of current.
MAX Sets to the maximum value of current.
SEND> LEV:CURR 1m<NL>//Set to 1mA. The A
unit suffix cannot be used.
LEVel:CURRent?
or CURRent:LEVel?
<NR3>
NR3 floating point
SEND> CURR?<NL>
RET>
1.000e+00<NL>
The A unit suffix cannot be used.
This command CANNOT be used in LIST MEAS page
and CORRECTION page.
LEVel:SRESistance {30,50,100}
VOLTage:SRESistance {30,50,100}
{30,50,100}
Where,
30
Sets the output impedance to 30
50
Sets the output impedance to 50
100
Sets the output impedance to 100
SEND> LEV:SRES 30<NL>//Set to 30 , the unit
cannot be added.
VOLTage:SRES?
or LEVel:SRES?
<NR1>
NR1 integer
SEND> LEV:SRES?<NL>
RET>
30<NL>