The CORRection:SPOT:FREQuency command sets the frequency point for
the specified frequency point correction.
This command executes the OPEN correction data measure for the specified
frequency correction.
This command executes the SHORT correction data measure for the
specified frequency correction.
122
1.800.561.8187
Query Syntax

CORRection:SPOT:STATe?

Query Response
{on,off}
Example
SEND> CORR:SPOT:STATe?<NL>
RET>
Command
CORRection:SPOT:FREQuendy <value>
Syntax
Parameter
value
A suffix multiplier can be used with this command. But
Example
SEND> CORR:SPOT:FREQ 100<NL>
SEND> CORR:SPOT:FREQ 10k<NL>
Query Syntax

CORRection:SPOT:FREQuency?

Query Response
<NR3>
Example
SEND> CORR:SPOT:FREQ?<NL>
RET>
Command

CORRection:SPOT:OPEN

Syntax
Example
SEND> CORR:SPOT:OPEN<NL>
RET>
Command

CORRection:SPOT:SHORt

Syntax
Example
SEND> CORR:SPOT:SHOR<NL>
RET>
www.
on<NL>
NR1,NR2 or NR3:Frequecny value.
1.000000e+03<NL>
pass/fail<NL> //Finish or Undone
pass/fail<NL> //Finish or Undone