The COMParator:TOLerance:NOMinal command sets the nominal value for
the tolerance mode of the comparator function.
The COMParator:TOLerance:BIN command sets the low/high limit values
of each BIN for the comparator function tolerance mode.
The COMParator:SLIM or COMParator:secondary command sets the
LOW/HIGH limit values for the secondary parameter.
1.800.561.8187
Command
COMParator:TOLerance:NOMinal <value>
Syntax
Parameter
Where,<value> is:
NR1, NR2 or NR3
A suffix multiplier can be used with this command. But
the suffix unit F/
Example
SEND> COMP:TOL:NOM 100N<NL>
SEND> COMP:TOL:NOM 1E-6<NL>
Query Syntax

COMParator:TOLerance:NOMinal?

Query Response
<NR3>
Example
SEND> COMP:TOL:NOM?<NL>
RET>
Command
COMParator:TOLerance:BIN <n>,<low limit>,<high
Syntax
limit>
Parameter
Where,<n>,<low limit>,<high limit> is:
n
low limit NR1,NR2 or NR3: low limit value
high limit NR1,NR2 or NR3: high limit value
Example
SEND> COMP:TOL:BIN 1,100P,200P<NL>
SEND> COMP:TOL:BIN 2,200E-6,300E-6<NL>
Query Syntax
COMParator:TOLerance:BIN? <n>
Parameter
Where,<n> is:
NR1 (1 to 9): Bin number
Query Response
<NR3:low limit>,<NR3:high limit>
Example
SEND> COMP:TOL:BIN? 2<NL>
RET>
Command
COMParator:SLIM <low value>,<high value>
Syntax
COMParator:secondary <low value>,<high value>
Parameter
Where, <low value>,<high value> is:
<low value>
<high value>
www.
1.00000e-06<NL>
NR1 (1 to 9): Bin number
1.00000e-06,2.00000E-6<NL>
NR1,NR2 or NR3: low limit value
NR1,NR2 or NR3: high limit value