Fetch:list - Teledyne T3LCR1002 User Manual

Precision lcr meters
Hide thumbs Also See for T3LCR1002:
Table of Contents

Advertisement

The FETCh:LIST? query sets the latest LIST measurement data of the
primary parameters, secondary parameters and comparator results into the
output buffer. Only applicable when in the [LIST MEAS] page view.
112
1.800.561.8187
Query Syntax

FETCh:LIST?

Query Response
<spot no>,<NR3:primary value>,<NR3:secondary
value>,<CMP result> // spot no:01-10
Example
SEND> FETC:LIST?<NL>
RET>
12,+3.27673e+00,L,02,+7.11030e-12,+3.48450e-
01,P,03,+7.11322e-12,+5.14944e-02,H,04,-
1.00000e+20,-1.00000e+20,-,05,-1.00000e+20,-
1.00000e+20,-,06,-1.00000e+20,-1.00000e+20,-,07,-
1.00000e+20,-1.00000e+20,-,08,-1.00000e+20,-
1.00000e+20,-,09,-1.00000e+20,-1.00000e+20,-,10,-
1.00000e+20,-1.00000e+20,-<NL>
(-1.00000e+20: The STEP is OFF)
Query Syntax
FETCh:LIST? <spot no>
Query Response
<spot no>,<NR3:primary value>,<NR3:secondary
value>,<CMP result>
Example
SEND> FETC:LIST? 2<NL>
RET>
(-1.00000e+20: The STEP is OFF)
www.
01,-2.98524e-
02,+7.11030e-12,+3.48450e-01,P<NL>
.com
Command Reference
information@itm.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

T3lcr1100T3lcr1300Everywhereyoulook t3lcr1002Everywhereyoulook t3lcr1100Everywhereyoulook t3lcr1300

Table of Contents