General Start-up

Function check of the emergency stop switches

Connect the system to the voltage and boot it.
Click "Emergency" for the patient registration.
Operate the EMERGENCY STOP switch on the C-arm.
The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop has been pressed"
Disengage the EMERGENCY STOP switch.
The safety switch moves audibly and the following message appears on the lower left
monitor edge: "The emergency stop switch has been released."
Document the function of the EMERGENCY STOP switch in the "Test protocol, C-arm
movement and emergency stop.
Siemens AG
Medical Solutions
SPR2-310.815.01.07.02
12.06
Page 25 of 64