10.5

Errors

Fault message
Description
Invalid device
No analyzer module was detected. At least one analyzer
configuration
module is required to use the device.
Firmware
The firmware of the basic device and the analyzer modules
versions
must be compatible with one another.
incompatible
Unknown
An analyzer module has been detected which is not
module type
supported in the previous firmware version.
detected
Memory error
A hardware fault, faulty firmware or no firmware at all has
detected
been detected in the following analyzer modules:
(e.g.) AM1
Possible causes:
No firmware is present on the analyzer module
The existing firmware is invalid
The program memory is faulty
In this case, the analyzer module starts in bootloader mode
and is ready to perform a firmware update.
If the error still exists despite the firmware update, the
analyzer module PCB must be replaced.
Cable fault
The installed analyzer modules could not be recognized
detected
correctly.
Possible causes: Faulty or incorrectly connected CAN bus
cable.
See also
Technical support (Page 76)
Rack-mounted device
Operating Instructions, 01/2013, A5E31874006-01
Maintenance and service
Actions
Install at least one analyzer module.
Please contact the service department.
Please contact the service department.
Please contact the service department.
Check the CAN bus connections.
10.5 Errors
61