Reinstalling The Fixed Platform; Errors And Corrective Actions - Siemens Clinitek Advantus Operator's Manual

Hide thumbs Also See for Clinitek Advantus:
Table of Contents

Advertisement

Reinstalling the Fixed Platform

If you turn the analyzer off during a run, or at any screen other than the Ready/
Run screen, the moving table may not be in its lowest position. If y ou then
remove the fixed platform, the moving table is pulled out at the same time. You
cannot reinstall the fixed platform because the pins of the moving table are in
the way.
Use the following procedure to resolve this problem:
1. Turn analyzer power on.
2. Let the analyzer initialize.
 An error displays because the fixed platform is not in place, but the moving
table is rotated into the correct position.
3. Turn analyzer power off.
4. Install the fixed platform. Refer to Performing the Daily Cleaning  ‚ page 37.
5. Turn analyzer power on.

Errors and Corrective Actions

Symptom
Changes made in
Setup are not
saved
Display is blank
Clinitek Advantus Operator's Guide
Possible Cause
Remedy  
You did not select
 Always select
Return to Ready/
after making changes in Setup.
after making
Run
changes.
No power
1. Listen for the fan.
2. If it is not running, turn analyzer 
3. Check that the power cord is
4. Turn analyzer power on.
Defective
Contact your local technical support
analyzer 
provider.
electronics
Return to Ready/Run
power off.
firmly connected to the analyzer 
and into a live AC electrical
outlet.
r   o
 u
 b 
l  
 e
 s
 o
 o
 t   i  
n
 g
57  