Figure 7-1 Electrical Connection - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Connecting
7.2 Electrical connection of the power supply
1. Connect the wires of the connecting cable to the appliance plug
Note
Tightening the screws
Tighten the screws in the appliance plug with a torque of
• 0.5 Nm to the connecting terminals,
• 0.3 Nm to the cable terminals (strain relief) and the connector housing.
These values apply to the optionally ordered power connector, see Spare parts list for
rack-mounted housing (Page 73).
2. Connect the appliance plug
40
power supply and PE conductor connections are correct!
3
2
1
Power supply unit
Appliance plug
Figure 7-1
Electrical connection
to the power supply unit
Connection of equipotential bonding
Operating Instructions, 01/2013, A5E31874006-01
. Ensure that the
.