Siemens SIPROCESS GA700 Operating Instructions Manual page 53

Continuous gas analysis
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Accompanying gas (concentration 100 vol. %)
Inert gases
Helium
Neon
Argon
Krypton
Xenon
Inorganic gases
Ammonia
Hydrogen bromide
Chlorine
Hydrogen chloride
Dinitrogen monoxide
Hydrogen fluoride
Hydrogen iodide
Carbon dioxide
Carbon monoxide
oxygen
Nitric oxide
Nitrogen
Nitrogen dioxide
Sulfur dioxide
Sulfur hexafluoride
Hydrogen sulfide
Water
Hydrogen
Zero point error due to diamagnetism or paramagnetism of some accompanying gases with nitrogen
Rack-mounted device
Operating Instructions, 01/2013, A5E31874006-01
as the reference gas at 60 °C and 1000 hPa absolute (according to IEC 1207/3)
Zero point deviation in vol. % O
absolute
He
Ne
Ar
Kr
Xe
NH
3
HBr
Cl
2
HCl
N
O
2
HF
HI
CO
2
CO
O
2
NO
N
2
NO
2
SO
2
SF
6
H
S
2
H
O
2
H
2
Commissioning
9.2 OXYMAT 7
2
+0.33
+0.17
-0.25
-0.55
-1.05
-0.20
-0.76
-0.94
-0.35
-0.23
+0.10
-1.19
-0.30
+0.07
+100
+42.94
0.00
+20.00
-0.20
-1.05
-0.44
-0.03
+0.26
53