2 - 6
Ready light indicator
Fig. 4

Check the indicators and control buttons

kV and mAs indication, technique factors

Switch the main switch to position "B" (battery power).
Select kV or mAs values by pressing the appropriate buttons. These values are indicated
at the display. When changing the kV values, the "Ready" light indicator sometimes goes
out while the generator adjusts to the new kV values.
NOTE

Radiation indicators

Select 81 kV and 10 mAs.
Make an exposure and confirm that the X-ray control console provides a visual indication
whenever X-rays are produced and that a signal audible to the operator is present
indicating that the exposure has been terminated.

Manual termination of exposure

The operator shall be able to terminate the exposure at any time during an exposure.
Select: 70 kV, 160 mAs
Make an exposure and immediately interrupt the exposure with the exposure release
switch. A message, "USE 20", appears on the display indicating that the exposure
release switch was released during exposure. It will also be indicated by 10 short beeps.
The message can be acknowledged by pressing the kV + button.
MOBILETT Plus HP
Radiation indicator
Pressing the button briefly increases/decreases the exposure
value by one step. Holding the button depressed results in a con-
tinuous increase/decrease.
SPR8-220.033.03
Rev. 02
Installation and setting
Page 6 of 12
10.02
SPS-UD