Checks with high voltage
Preparation
This section describes how to check:
•
Radiographic voltage.
•
Tube current.
•
mAs values.
CAUTION!
Radiation
NOTE!
Check that the high-voltage plug is plugged into H1 and locked.
kV>17
kV>50
WR_AUS
MASOLL
MAIST
IHSOLL
IHIST
UZIST
ISCHWING
KVSOLL
KVIST
0VA
KVE
KVA
ZBL
BRAKE
0VD
Figure 1 P.C. board D705
1
Set the switch S2 (SS) to "ON" (upper position) on p.c. board D702.
2
Remove the jumper and connect the mAs meter to mAs measuring sockets X3 and
X4 on p.c. board D710.
Siemens-Elema AB
Solna, Sweden
V39
V41
V40
X706
X707
X705
X704
X713
X710
X712
X711
X708
X714
X715
X716
X717
X709
R136
SPS-UD
RX B7-120.033.01.10.02
X703
05.98
11-1
MAMMOMAT 300
65 02 186