Semiconductor Survey "Bellows" Probe
Assembly (Option) - Part No. 42444
The dust filter assembly can be removed, as shown in Fig.
5-6, and replaced with a water filter assembly. The water
filter assembly is available as an accessory. Replace the
probe sealing washer if it shows signs of damage or wear.
Probe Sealing Washer
Fig. 5-6 Semiconductor Survey "Bellows" Probe
Filter Assembly
Caution: The instrument should never be switched on
without suitable filters installed.
If a blockage occurs the 'SAMPLE FAULT' indicator is
displayed. Check the sample line and filter assembly for
blockage. Press
(Button 1) to clear the 'SAMPLE
FAULT' message. Replace the filter if the message does
not clear.
5-6