Teledyne GMI is prohibited. Reverse engineering is not permitted. LIABILITY Every care has been taken in the preparation of this document, but Teledyne GMI do not accept any responsibility for errors or omissions and their consequences. Information in this document is subject to change without notice. This document does not constitute a specification or basis for a contract.
2014/34/EU. Any right of claim relating to product liability or consequential damage to any third party against Teledyne GMI is removed if the warnings are not observed. AREAS OF USE Exposure to certain chemicals can result in a loss of sensitivity of the flammable sensor.
CONTENTS COPYRIGHT ..............i LIABILITY ...............i MODIFICATION NOTICES ..........i SOFTWARE ..............i DISPOSAL ADVICE ............i SAFETY ................. ii AREAS OF USE ............ii STORAGE, HANDLING AND TRANSIT ......ii INTRODUCTION ........1-1 GENERAL INFORMATION ......2-1 Ranges of Operation..........2-1 EXT Range (External Semiconductor Sensor), 0 –...
Page 6
FIRST RESPONDER CG - USER HANDBOOK Carbon Monoxide, 0 – 1000 ppm (CSM mode only) ............ 2-4 Hydrogen Sulphide, 0 – 100 ppm (CSM mode only) ............ 2-5 Oxygen, 0 – 25% (CSM and PURGE modes only) ......2-6 Alarms................ 2-7 General Information ..........
Page 7
CONTENTS OPERATION ..........3-1 Instrument Features........... 3-1 Button Panel .............. 3-2 CGI mode ..............3-2 Switching ON in CGI Mode ........3-3 Calibration Date Features ........3-4 Switching OFF ............3-4 Switching Pump OFF / ON ........3-4 Manual Datalog ............3-4 Notes on CGI mode: ..........
Page 8
FIRST RESPONDER CG - USER HANDBOOK Switching Pump OFF / ON ........3-12 Zero Vol Gas Range ..........3-12 Manual Datalog ............. 3-13 Notes on PURGE mode: ........3-13 Summary of PURGE Button Operation ....3-13 TRAC mode ............. 3-14 Switching ON in TRAC Mode ........
Page 10
FIRST RESPONDER CG - USER HANDBOOK Display ..............A-2 Sampling System ............ A-2 Power Source ............A-3 FIELD CALIBRATION ......... B-1 Selectable Ranges in FCM ........B-2 Entering FCM............. B-3 Zeroing the Instrument..........B-4 Field Calibration Procedure ........B-5 Quitting FCM.............. B-8 Quit and Save Changes ..........
Oxygen (O2), for internal atmosphere monitoring where odour call response is required. The First Responder CG also features a Parts Per Million (PPM) range to increase sensitivity to smaller leaks. This is achieved using the external (EXT) semiconductor sensor located at the end of the ‘flexi’...
Page 12
The instrument monitors all gases continuously, providing all round protection for users. The First Responder CG is equipped with both audible and visual alarms and Geiger indications on the EXT flammable range. The First Responder CG includes the following ranges: •...
Page 13
INTRODUCTION The First Responder CG is supplied in a carrying case with a selection of accessories. For a comprehensive list of accessories supplied with the instrument, and additional accessories available, see Section 7 ACCESSORIES. The main features of the instrument are: •...
GENERAL INFORMATION Ranges of Operation The instrument calibration gas is shown on the instrument service label. EXT Range (External Semiconductor Sensor), 0 – 4,500 ppm - 20 %LEL Flammable (TRAC mode only) This range displays flammable gas parts per million (ppm) content up to 4,500pmm, then autoranges to %LEL (concentration up to 20% LEL).
FIRST RESPONDER CG - USER HANDBOOK Fig. 2-1 EXTernal Range PPM, 0 – 4500 ppm (CGI mode only) This range will measure very low levels of gas and indicates the PPM concentration of gas in air (45 ppm is equivalent to 0.1% LEL).
GENERAL INFORMATION LEL, 0 – 100% (CGI and CSM modes only) The LEL range indicates the explosibility of the flammable gas in the sample. This is displayed as a percentage of the lower explosive limit (LEL) of the gas. For methane 100% LEL corresponds to 4.5% Volume methane in Air.
FIRST RESPONDER CG - USER HANDBOOK Volume Gas, 0 – 100 % (CGI, CSM and PURGE modes only) This range displays the total volume of flammable gas with respect to air. When the Volume Gas range is selected, GAS is displayed in the top right corner of the LCD.
GENERAL INFORMATION Fig. 2-5 Carbon Monoxide Range Hydrogen Sulphide, 0 – 100 ppm (CSM mode only) The Hydrogen Sulphide range measures on the ppm scale. The digital display resolves to 1 ppm H2S. The analogue bar graph follows in 4 ppm steps. Above 100 ppm, EEEE is displayed.
FIRST RESPONDER CG - USER HANDBOOK Oxygen, 0 – 25% (CSM and PURGE modes only) This range displays the percentage Oxygen content of the sample. The instrument range is displayed in the top right corner of the LCD. The digital display resolves the signal to 0.1% O...
GENERAL INFORMATION Alarms General Information Alarms can have both audible and visual indications active. The audible alarm is rated up to 85 dB(A) at 1ft. (0.3m) approx. The visual alarm is a red LED indicator which protrudes from the instrument top plate, allowing viewing from any angle above the top plate.
‘EXT’ range. This means that, as the gas concentration increases, both the sounder and the red LED pulse at an increasing rate. Datalogging The First Responder CG provides the user with an automatic and manual datalogging function. Automatic Datalogging By default, the instrument has automatic datalogging enabled in CSM and PURGE modes and is confirmed by the ‘STORE’...
GENERAL INFORMATION Fig. 2-8 Automatic Datalogging Enabled During automatic datalogging the instrument stores the values of all gas ranges together with the current time and date at 60 second intervals. Manual Datalogging By default, the instrument has manual datalogging enabled in all modes (CGI, CSM, PURGE and TRAC).
Filters Probe Handle Assembly - Part No. 12481 A number of different filter types are available from Teledyne GMI. The minimum requirement is a cotton particulate filter and a hydrophobic filter. These filters, incorporated in the probe handle assembly, are available from Teledyne GMI.
A replacement Dust Filter disc, which is incorporated in the Semiconductor Survey “Bellows” Probe, is available from Teledyne GMI. A Water Filter assembly is available as an accessory for use with this probe. Filters should be checked periodically and, where appropriate, changed to ensure a clean sample path.
FIRST RESPONDER CG - USER HANDBOOK Liquid Crystal Display (LCD) The LCD shows the current gas readings in both analogue and digital form together with operational and status information. The display is protected by a toughened glass cover. Backlighting is provided to enable the display to be seen under low ambient light conditions.
OPERATION Instrument Features The First Responder CG front plate features a panel of four operating buttons, a backlit LCD screen incorporating the analogue bar graph, a visual alarm LED and an infrared communication port. Button Backlit LCD Infrared Panel Screen...
Fig. 3-2 Button Panel Button function is dependent on the operating mode of the instrument. The First Responder CG has four operating modes, CGI, CSM, PURGE and TRAC, each accessed by switching the instrument ON with buttons 1, 2, 3 and 4 respectively.
OPERATION Switching ON in CGI Mode Pressing and holding (Button 1) turns the instrument ON in CGI mode and switches ON the pump. This initiates the instrument’s warm-up cycle, an example of which is shown in Figure 3-3. Pump Test (user to block the inlet for PASS) Fig.
FIRST RESPONDER CG - USER HANDBOOK Calibration Date Features At the end of warm-up and before the First Responder CG is ready for measuring, the instrument will indicate on the display, how many days remain until calibration is required, as shown in Figure 3-4:...
OPERATION Notes on CGI mode: • This mode has PPM autoranging to LEL, and, LEL autoranging to Vol Gas. • The pump can be switched OFF. • Manual datalogging is enabled. Summary of CGI Button Operation Double Press and Single Press Press Hold Button 1...
FIRST RESPONDER CG - USER HANDBOOK CSM mode (Confined Space Monitor) In this mode the instrument operates as a safety monitor for use when entering confined spaces which may contain hazardous gas mixtures (flammable, toxic and/or asphyxiant). The instrument samples the ambient atmosphere to which...
(user to block the inlet for PASS) Fig. 3-6 CSM Warm-up Note: STEL and LTEL alarms not illustrated Calibration Date Features At the end of warm-up and before the First Responder CG is ready for measuring, the instrument will indicate on the...
FIRST RESPONDER CG - USER HANDBOOK display, how many days remain until calibration is required, as shown in Figure 3-7: Calibration Due (Alternate flash) in 25 days Figure 3-7 Days until Calibration Required If calibration is overdue, the instrument will automatically switch OFF.
OPERATION a double press of (Button 4) will acknowledge each alarm in order of priority. Notes on CSM mode: • This mode has LEL, Vol GAS, O2, CO and H2S ranges. • Audible and Visual alarms are only active in the LEL, O2, CO and H2S ranges.
FIRST RESPONDER CG - USER HANDBOOK PURGE mode This mode facilitates the purging of pipework / installations and measures Volume Methane and Oxygen only. There are no alarms or confidence signal in this mode. Switching ON in PURGE Mode Pressing and holding (Button 3) turns the instrument ON in PURGE mode and switches ON the pump.
(user to block the inlet for PASS) Fig. 3-9 PURGE Warm-up Calibration Date Features At the end of warm-up and before the First Responder CG is ready for measuring, the instrument will indicate on the display, how many days remain until calibration is required,...
FIRST RESPONDER CG - USER HANDBOOK Calibration Due (Alternate flash) in 25 days Figure 3-10 Days until Calibration Required If calibration is overdue, the instrument will automatically switch OFF. The pre-set re-calibration interval is 30 days. This period can be altered as an option, however, you should ensure that the instrument is always within its calibration period prior to use.
OPERATION Note: Negative Volume Gas readings are displayed to a minus value of -10%, thereafter, ‘EEE’ is displayed. Manual Datalog Manual datalogging is enabled in this mode. To perform a manual datalog, a double press of (Button 3) is required. Notes on PURGE mode: •...
FIRST RESPONDER CG - USER HANDBOOK TRAC mode (Leak Detection) In this mode the instrument utilises a semiconductor sensor for near instantaneous response to enable rapid leak detection of PPM Methane in small fittings. Switching ON in TRAC Mode Pressing and holding (Button 4) turns the instrument ON in TRAC mode.
(user to block the inlet for PASS) Fig. 3-12 TRAC Warm-up Calibration Date Features At the end of warm-up and before the First Responder CG is ready for measuring, the instrument will indicate on the display, how many days remain until calibration is required,...
FIRST RESPONDER CG - USER HANDBOOK Calibration Due (Alternate flash) in 25 days Figure 3-13 Calibration Due Date If calibration is overdue, the instrument will automatically switch OFF. The pre-set re-calibration interval is 30 days. This period can be altered as an option, however, you should ensure that the instrument is always within its calibration period prior to use.
OPERATION Notes on TRAC mode: • This mode only has EXT range (ppm to %LEL). • By default, only the visual geiger alarm is enabled. The audible geiger alarm can be toggled by a single press of (Button 4). Display will alternate ‘EXT’ and ‘Aud’. •...
The instrument will now switch to the selected mode. Fig. 3-15 Instrument Switching OFF Connection of Probe The First Responder CG instrument is equipped to accept two standard probe input types. CGI, CSM and Purge modes Only A Standard Probe and Handle assembly which is connected to the instrument via sample line inlet.
Page 45
OPERATION TRAC Mode Only The EXT range Semiconductor Sensor “Flexi” Probe provides fast response gas detection and is connected to the instrument via an 8-Pin Plug. Flexible Probe 8-Pin instrument connector plug Probe Sensor Fig. 3-17 Semiconductor Sensor “Flexi” Probe An optional Semiconductor Survey “Bellows”...
FIRST RESPONDER CG - USER HANDBOOK Instrument connection points for the three types of probe are shown in Figure 3-19. Connection for Connection for Probe Sample Line 8-Pin Probe Connector Plug Fig. 3-19 Probe Connection Operator Messages / Fault Flags Various messages can appear on the LCD screen to indicate instrument status.
Page 47
OPERATION ‘SAMPLE FAULT’ This indicates a problem with the instrument’s flow due to the sample path being blocked, water ingress, a blocked filter or pump failure. The pump will switch off. The sample path should be checked and Button One pressed to clear sample fault and re-start the pump.
Page 48
FIRST RESPONDER CG - USER HANDBOOK 3-22...
PROBES The First Responder CG instrument is equipped to accept the following probes and associated accessories: Standard Probe and Handle Assembly Part No. Description 12393 31ins. (80cm.) Plastic Probe - Solid End 12481 Probe Handle Assembly - incl. filters 12712 Clear Sample Line x 4ft 10ins.
PROBES Semiconductor (EXT) Survey Probe (Optional) Part No. Description 42444 Semiconductor Survey (Bellows) Probe Assembly incl. Dust Filter 42416 Water Filter Assembly Semiconductor Survey Probe c/w Dust Filter Sample Line 8-Pin DIN Plug Water Filter Assembly Fig. 4-3 Instrument / Semiconductor Survey (Bellows) Probe For a comprehensive list of accessories supplied with the instrument, and additional accessories available, see Section...
Replacing Alkaline (LR20) Batteries All four batteries should be replaced at any one time and in a safe area. Teledyne GMI only recommend the use of Energizer or Duracell cells. 1. Using the Hex Driver supplied (GMI Part No. 12451), loosen the two instrument base screws.
Page 54
FIRST RESPONDER CG - USER HANDBOOK 2. Remove battery cover. Fig. 5-2 Remove Battery Cover 3. Remove the old batteries. 4. Check battery compartment for damage to spring contacts or corrosion on springs. C a u t i o n : U n d e r n o c i r c u m s t a n c e s s h o u l d rechargeable batteries be substituted for alkaline.
OPERATOR MAINTENANCE Probe Filter Replacement Probe Handle Assembly - Part No. 12481 Hydrophobic and cotton particulate filters in the probe handle minimise the danger of water and dust ingress. Caution: The instrument should never be switched on without suitable filters installed. If a blockage occurs, the ‘SAMPLE FAULT’...
FIRST RESPONDER CG - USER HANDBOOK 2. Remove the cotton particulate filter and discard. 3. Remove the hydrophobic filter. 4. Clean the probe handle to make sure that it is free from dirt and water. 5. Fit a new cotton particulate filter.
Page 57
OPERATOR MAINTENANCE To replace the filter disc, proceed as follows: 1. The yellow end cap is a push fit on to the probe detector cap. Carefully pull then remove the end cap. Note: If the cap is a tight fit, grip the cap then remove using a twisting motion.
FIRST RESPONDER CG - USER HANDBOOK Semiconductor Survey “Bellows” Probe Assembly (Option) - Part No. 42444 The dust filter assembly can be removed, as shown in Fig. 5-6, and replaced with a water filter assembly. The water filter assembly is available as an accessory. Replace the probe sealing washer if it shows signs of damage or wear.
Page 59
OPERATOR MAINTENANCE Dust Filter (Accessory) To replace the Dust Filter in the Dust Filter assembly, proceed as follows: 1. Unscrew the filter assembly (Fig. 5-7). Dust Filter Housing Assembly Dust Filter Probe Washer Fig. 5-7 Dust Filter Assembly 2. Remove Dust Filter, then discard. 3.
Page 60
FIRST RESPONDER CG - USER HANDBOOK Water Filter (Accessory) To replace the Hydrophobic Filter in the Water Filter assembly, proceed as follows: 1. Unscrew the filter assembly (Fig. 5-8). Water Filter Housing Assembly Hydrophobic Filter Assembly Probe Washer Fig. 5-8 Water Filter Assembly 2.
OPERATOR MAINTENANCE Probe Handle Assembly (Option) Part No. 13561 Fitting of an In-line filter assembly to the probe handle, as shown in Figure 5-9, will minimise the danger of water and / or dust ingress. Fig. 5-9 Probe Handle and In-line Filter Assembly Caution: The instrument should never be switched on without suitable filters installed.
Page 62
FIRST RESPONDER CG - USER HANDBOOK In-line Dust Filter (Accessory) To replace the Dust Filter in the Dust Filter Housing assembly, proceed as follows: 1. Unscrew the filter housing assembly (Figure 5-10). Probe Sealing Washer Dust Filter Probe Washer Dust Filter Housing Assembly Fig.
Page 63
OPERATOR MAINTENANCE In-line Water Filter (Accessory) To replace the Hydrophobic Filter in the Water Filter Housing assembly, proceed as follows: 1. Unscrew the filter housing assembly (Fig. 5-11). Water Filter Housing Assembly Hydrophobic Filter Probe Sealing Washer yellow Probe Washer label Fig.
FIRST RESPONDER CG - USER HANDBOOK Flue Probe Coalescing Filter Assembly (Accessory) - Part No. 42215 A coalescing filter is fitted in the filter bowl and housing assembly to minimise the danger of water ingress. Caution: The instrument should never be switched on without suitable filters installed.
Page 65
OPERATOR MAINTENANCE 3. Using a dry cloth, clean the bowl, housing and spindle to make sure that they are free from dirt and water. 4. Fit a new coalescing filter to the spindle then replace in filter housing. 5. Replace the bowl and tighten to secure. 5-13...
Page 66
FIRST RESPONDER CG - USER HANDBOOK 5-14...
CALIBRATION The instrument has been calibrated for a particular flammable gas mixture. Where any doubt exists the instrument should be returned to Teledyne GMI or an authorised distributor for calibration. Four methods of calibration are possible: • Automatic Calibration. The GMI Gas Delivery Unit (GDUnet) allows full automatic calibration.
12 month period can be expected. This is no guarantee, however, as the precise application of the product is unknown to Teledyne GMI. Individual codes of practice may dictate shorter periods. Regular checking establishes a pattern of reliability and enables the calibration check period to be modified in line with operational experience.
ACCESSORIES Accessories Supplied with First Responder CG Instrument Part Number: 44542SCG are as follows: Part Number Description 42710 Carrying Case 10278 Alkaline Battery (LR20) x 4 12528 Carrying Harness 12451 Hex Driver (for base screws) 42200 Semiconductor Sensor (Flexi) Probe...
Page 70
Probe Sealing Washer (use with 42183, 42184, 42444 or 42416) 42235 Probe Washer (use with 42183, 42184, 42444 or 42416) Note: For other sampling probes and accessories, and for calibration gases, contact Teledyne GMI directly or a Tele- dyne GMI approved distributor.
ADDITIONAL INFORMATION Training Training Courses are available on all GMI products. Contact GMI Marketing Department for further details: Tel: +44 (0) 141 812 3211 e-mail: gmi_sales@teledyne.com Website www.teledynegasandflamedetection.com...
FIRST RESPONDER CG - USER HANDBOOK Size 7.08” (180mm.) x 3.74” (95mm.) x 4.13” (105mm.) Weight 3.75lbs. (1.7kg) with alkaline batteries Operating Temperature F to 122 F (-20 C to 50 Humidity 0 – 95% RH Construction Moulded polypropylene case protected to IP54...
TYPICAL OPERATING PARAMETERS Semiconductor Survey (Bellows) Probe connected to instrument with electrical harness and sample line and uses instrument integral pump with flow fail sensor. The sample path is protected by a dust or hydrophobic filter and automatic pump switch off. Power Source 4 ‘D’...
Field calibration allows simple calibration to be carried out in the field without the use of additional test equipment. Other calibration procedures require the use of the Teledyne GMI Manual Calibration software or the Workshop System. In Field Calibration Mode (FCM) the buttons perform the functions indicated in Figure B-1.
FIRST RESPONDER CG - USER HANDBOOK To simplify button operation when calibrating the instrument, an overlay card, shown in Figure B-2, is available and can be placed over the top face of the instrument to identify calibration button functions. Contact Teledyne GMI for details.
FIELD CALIBRATION Entering FCM 1. Switch the instrument on and allow it to complete its warm-up checks. 2. Double press (Button 1) to initiate instrument switch off. While OFF is displayed in the LCD and before the instrument actually switches off, enter the four button access code.
FIRST RESPONDER CG - USER HANDBOOK Press and Single Press Double Press Hold Button 1 Toggle Pump Exit CAL Mode Enter SPAN Mode On / Off Button 2 Zero Current Save CAL Data Next Range Range Button 3 Increase Exit SPAN –...
FIELD CALIBRATION Field Calibration Procedure 1. Zero gas range before attempting calibration. See previous section ZEROING THE INSTRUMENT for details. 2. Make sure that the instrument pump is running and the gas range selected is compatible with the calibration gas. Note: A single press of (Button 1) toggles the pump OFF / ON.
Page 82
FIRST RESPONDER CG - USER HANDBOOK 4. Turn the regulator valve counter clockwise to open the valve slightly. Make sure that the gas is flowing before connecting the sample tubing to the instrument, otherwise an instrument sample fault may occur.
Page 83
FIELD CALIBRATION a. A single press of (Button 3) will produce small incremental changes to increase display reading, or a single press of (Button 4) will produce small decremental changes to decrease display reading, until the displayed gas value corresponds to the concentration of the calibration gas.
FIRST RESPONDER CG - USER HANDBOOK 11. Make sure that the regulator valve is in the fully closed position (Off) then disconnect the regulator from the gas cylinder (turn regulator body in a counter clockwise direction). 12. Replace the cap on the calibration gas cylinder.
APPLICATIONS Plastic (Solid End) Probe This probe is supplied with the First Responder CG and is designed for use with barhole probing and below ground surface gas monitoring. Fig. C-1 Solid End Probe The solid end feature prevents blocking of the probe when...
Page 86
FIRST RESPONDER CG - USER HANDBOOK The Plastic Probe (Part No. 12393) is generally used with the Probe Handle Assembly (Part No. 12481), incorporating both Cotton and Hydrophobic Filters to prevent the ingress of water or dirt. The Sample Line (Part No. 12712) connects the probe handle to the instrument and allows the pump to draw the sample into the instrument analysing chamber.
APPLICATIONS Semiconductor Sensor (Flexi) Probe This external (EXT) range diffusion probe (Part No. 42200) is designed to provide a faster speed of response in detecting small fitting leaks, by eliminating the delay time taken by the pump drawing the sample into the instrument analysing chamber.
Page 88
FIRST RESPONDER CG - USER HANDBOOK Fig. C-3 Semiconductor Sensor (Flexi) Probe Applications...
Page 89
APPLICATIONS The Semiconductor Probe assembly has a carry pouch, which is available as an accessory, to provide protection during transit. The Probe Sensor is protected by a removable Plastic End Cap (Part No. 42187). The cap incorporates a Filter Disc (Part No. 42197) to prevent the ingress of dust. The Probe is connected to the instrument via an 8-pin DIN plug.
FIRST RESPONDER CG - USER HANDBOOK Semiconductor Survey (Bellows) Probe This probe is designed for use with the First Responder 2 and is available as an accessory. This external (EXT) range diffusion probe (Part No. 42444), complete with filter and bellows fitting, is designed to provide a near instantaneous response in detecting general survey leaks.
Page 91
APPLICATIONS The probe incorporates a Dust Filter, with replaceable Disc, to prevent the ingress of dust. A Water Filter assembly (Part No. 42416), complete with replaceable hydrophobic filter is also available as an accessory. The Probe is connected to the instrument via an 8-pin DIN plug and sample line.
FIRST RESPONDER CG - USER HANDBOOK Bellows Probe This probe is also available as an accessory. The Bellows Probe (Part No. 13563) provides a method of obtaining more consistent readings by reducing the effect of wind and air dilution. Fig. C-6 Bellows Probe The Bellows Probe is of stainless steel construction housed in a flexible rubber boot.
APPLICATIONS Note: This model of Probe Handle does not incorporate an in-line filter. A Water Filter Assembly (Part No. 42183) or Dust Filter Assembly (Part No. 42184) is available and can be fitted as an accessory, if required. Refer to Chapter 7 ACCESSORIES.
Page 94
FIRST RESPONDER CG - USER HANDBOOK The Swan Neck Probe (Part No. 13655) has a number of small balanced holes in the probe length to give increased coverage in surveys. To prevent damage to the probe and provide an unrestricted sample path, the probe is supplied with a shroud assembly incorporating two skids which prevent probe contact with the ground surface.
APPLICATIONS Stainless Steel (Flue) Probe This probe is also available as an accessory. Fig. C-8 Flue Probe When a flue sample is required, a Stainless Steel open ended probe (Part No. 13413) with Coalescing Filter Assembly (Part No. 42215), incorporating Coalescing Filter (Part No. 75174), is generally used.
Page 96
FIRST RESPONDER CG - USER HANDBOOK C-12...
INDEX Symbols CALIBRATION 6-1, B-1 1 3-21 Calibration Date Features 3-4, 3-7, 3-11, 3-15 Calibration Procedure B-5 ACCESSORIES 7-1 Calibration Validity 6-2 Accuracy A-1 Carbon Monoxide, 0 - 1000 Acknowledge Alarms 3-8 ppm 2-4 ADDITIONAL CGI Button Operation 3-5 INFORMATION 8-1 CGI mode 3-2 ADVICE, DISPOSAL i Changing Gas Range...
Page 98
FIRST RESPONDER CG - USER HANDBOOK DISPOSAL ADVICE i Dust Filter 5-7, 5-10 Gas A-1 GENERAL INFORMATION EEEE 3-21 Entering FCM B-3 EXT Range 3-15 Handle Assembly - Part No. EXT Range (External 13561 5-9 Semiconductor Sensor) Handle, Probe 2-10, 2-11...
Page 100
FIRST RESPONDER CG - USER HANDBOOK Semiconductor Survey (Bellows) Probe C-6 Quit And Save Changes Semiconductor Survey “Bellows” Probe Assembly Quiting FCM B-8 2-11, 5-6 Quit Without Saving Size A-2 Changes B-8 Small Plastic (Solid End) Probe C-1 SOFTWARE i...
Page 101
INDEX Temperature, Operating TRAC Button Operation 3-16 TRAC mode 3-14 Training 8-1 TRANSIT ii Underwriters Laboratories Use, Before 2-12 Validity, Calibration 6-2 Volume Gas, 0 - 100 % 2-4 Water Filter 5-8, 5-11 Weight A-2 World Wide Web 8-1 Zero A-1 ZERO, CHECK 3-20 Zero EXT Range 3-15 ZERO FAULT 3-20...