Maintenance
Pump Head Maintenance (Tool Free)
3
Turn the pump head assembly upside down.
5
Remove the inlet valve and the outlet valve from the
primary pump head.
NOTE
Clean the valves by sonication, if appropriate. A good
cleaning solution is 50 % isopropanol in water.
7
Place the two pump heads in the pump head holder.
1260 Infinity II Flexible Pump User Manual
4
Remove the link plate by gently pulling it off the pump
head assembly.
The two pump chambers are now isolated.
6
Binary/High-Speed Pumps only: Remove the
high-pressure filter from the secondary pump head.
8
Remove the pump head screws from the back of the
pump heads.