Bacharach 100 Operation & Maintenance Manual page 52

Combustion gas analyzer
Table of Contents

Advertisement

6. Connect the tubng of the Calbraton Kt to the GAS nlet of the an-
alyzer. Note that the gas fttng (Item 6) s only requred on Fyrte
Pro Models 105 and 125.
7. Adjust the regulator for approxmately 2 SCFH of excess flow (see
Fgure 5.1).
8. After the analyzer has stablzed (2 to 3 mnutes), use the  and
 buttons to set the dsplayed value to match the CO concentraton
stamped on the gas cylnder.
9. Press eNTeR to store the new calbraton values and move to the
next calbraton screen, or hold down eNTeR for 2 seconds to store
the new calbraton values and enter the Run Mode, or press the
I/O button to ext the Calbraton Mode and turn OFF the analyzer
wthout savng the changes.
10. Dsconnect tubng from analyzer and turn off gas flow.
46
Figure 5-1. Calibration Kit Hookup
Fyrite Pro
Instruction 24-9395

Advertisement

Table of Contents
loading

This manual is also suitable for:

100 printer kit105110120125125 printer kit ... Show all

Table of Contents