3.5.3 CO Channel
The CO Channel Setup Screen s labeled "CO."
1. Enter the Setup Mode per Secton 3.5.1. If necessary, repeatedly
press the eNTeR button untl "CO" s dsplayed.
2. Press the  or  button untl the desred parameter s dsplayed.
- no
Sensor dsabled. Use ths functon to turn off the CO chan-
nel f the sensor s mssng or known to be bad to avod nac-
curate CO readngs from appearng.
- 2Ero Manual Zero. When the CO channel s set to manual zero,
the analyzer does not zero the CO sensor to ambent cond-
tons durng start up.
- A2 Auto Zero. Wth the CO channel set to auto zero, the CO
sensor s zeroed to the ambent CO level durng start up.
Important: When using this mode the analyzer must
be turned ON in fresh air; otherwise, incorrect CO
readings will occur.
3. Press eNTeR to move to the next Setup Screen, or I/O to ext setup.
3.5.4 Draft Channel Setup
The Draft Setup Screen s labeled "PrES Unt".
1. Enter the Setup Mode per Secton 3.5.1. If necessary, repeatedly
press the eNTeR button untl "PrES" s dsplayed.
2. Press the  or  button untl the desred parameter s dsplayed.
- no
- WC
- mB
- HPa
3. Press eNTeR to move to the next Setup Screen, or I/O to ext setup.
Instruction 24-9395
Setup
Pressure sensor dsabled or not nstalled.
nches of Water Column
Mllbars
Hecto Pascals