Konica Minolta CM-36dGV Instruction Manual page 35

Hide thumbs Also See for CM-36dGV:
Table of Contents

Advertisement

Symptom
Displayed message
Install the gloss calibration
plate and appropriate target
mask, and then perform
calibration.
Light shielding during 0%
calibration was insufficient.
Light intensity at 100%
calibration was insufficient.
A target mask is not installed.
Install the target mask.
The UV control coefficient does
not exist. Perform UV control
using the UV Control Software.
Calibration data has not been
set.
Value exceeds the range
that can be measured by the
instrument.
The specified operation
cannot be executed because
measurement is in progress.
The specified operation cannot
be executed because motor
drive is in progress.
Possible cause
Gloss calibration was not
performed using the correct
procedure.
0% calibration was not
performed using the correct
procedure.
100% calibration was not
performed using the correct
procedure.
A target mask is not installed.
Measurement is not possible
because UV control has not
been performed for the set
conditions.
The calibration data for white
calibration has not been
written.
Color measurement of that
specimen is not possible.
The operation cannot
be executed because
measurement is in progress.
The operation cannot be
executed because the motor
that performs switching of the
measurement area and SCI/
SCE is currently operating.
33
Correction
Perform gloss calibration using
the provided gloss calibration
plate and appropriate target
mask.
Install the white calibration
plate correctly and perform
correct 0% calibration with
light completely blocked
between the illumination
window and receptor window.
Install the white calibration
plate correctly and perform
correct 100% calibration for
the specimen that will be
measured.
Install the appropriate target
mask before performing
measurement.
Perform UV control for the set
UV conditions from the UV
Control Software that is provided
with SpectraMagic NX before
performing measurement.
Write the calibration data to
the instrument.
To perform color measurement,
measure using a different
specimen.
Wait for measurement to
end before executing the
operation.
Wait for motor operation to
end before executing the
operation.

Advertisement

Table of Contents
loading

Table of Contents