Measurement Procedure

• This manual explains the measurement preparation procedure, specimen setting procedure, and other
procedures for measurement with the CM-36dGV.
• Control is performed and measurement is executed from the computer connected to the instrument by using
the optional SpectraMagic NX (CM-S100w) color data software.
• For the measurement procedure using SpectraMagic NX (CM-S100w), refer to the CM-S100w instruction manual.
Flow of Measurement
Connecting to the computer
Connecting the AC adapter
Starting the computer
(Start Windows.)
Launching the software
Power ON
Installing the target mask
Executing zero calibration
(0% calibration)
Executing white calibration
(100% calibration)
* When using WAA (Wavelength Analysis &
Adjustment), execute it after white calibration. WAA
execution requires approximately 20 seconds.
Executing gloss calibration (*)
Setting a specimen
Executing measurement
Power OFF
Use the USB cable to connect the instrument and computer. (P.18)
Use the AC adapter and connect the instrument to the outlet. (P.19)
Turn ON the power of the connected computer.
Launch SpectraMagic NX (CM-S100w) and enable control.
Turn the instrument power ON. (P.19)
<For reflectance measurement>
Install the target mask to use for
measurement. (P.20)
Install the zero calibration box
and execute zero calibration.
(P.22)
Install the white calibration
plate and execute white
calibration. (P.23)
Install the gloss calibration plate and
execute gloss calibration. (P.24)
Set the specimen into the instrument. (P.27)
Execute measurement from SpectraMagic NX (CM-S100w). Or set to
remote measurement mode and execute measurement by pressing the
measurement key on the instrument.
After measurement is completed, turn the instrument power OFF. (P.19)
Exit SpectraMagic NX (CM-S100w) and turn the computer power OFF.
17
<For transmittance measurement>
Install the target mask for LAV
(Ø25.4 mm) measurement and the
white calibration plate. (P.23)
Install the transmittance zero
calibration plate and execute
0% calibration. (P.25)
Execute 100% calibration using
an empty chamber. (P.26)