• Perform measurement when the specimen or the inside and outside of the container containing the specimen
are free of scratches, fingerprints, and other dirt. If the measurement surface is grasped with a hand,
fingerprints will be left on it. Hold the specimen at a part which is not the measurement surface.
• If there are bubbles on the inside of the container containing the specimen, accurate measurement will not be
possible. Remove the bubbles before performing measurement. (Bubbles may form spontaneously when a
specimen is left in the container.)
• If there is condensation or other water droplets on the specimen or the container containing the specimen,
accurate measurement will not be possible.
• When measuring a liquid which contains minute particles, accuracy will be unstable due to precipitation of the
particles.
3 Close the transmittance specimen chamber cover.
4 Use the optional SpectraMagic NX (CM-S100w) color
data software and perform measurement. When remote
measurement mode is selected, measurement can
also be performed using the measurement key on the
instrument status panel.
Haze Measurement
When the white calibration plate is installed at the specimen measuring port for reflectance measurement, the
measurement type is di: 0°. When the zero calibration box is installed instead of the white calibration plate, the
measurement type is de: 0°.
Although the illumination/viewing system does not completely match the definition of haze (ASTM D 1003), it can
be used as a relative value for management.
When using the optional SpectraMagic NX (CM-S100w) color data software for haze measurement, haze is
calculated from these 2 measurements.