Siemens ARCADIS Orbic Installation And Startup page 21

Hide thumbs Also See for ARCADIS Orbic:
Table of Contents

Advertisement

3D Reconstruction Upgrade Part 1
Attach the two limit switches so that the cam enables the switching mechanisms at the
switches
When set correctly, you can hear the click emitted by the limit switches.
Angulation 0°.
Test the functioning of the limit switches by moving the C-arm into the +95° und -95° or-
bital positions.
Check these settings at least twice.
Siemens AG
Medical Solutions
(Fig. 12 / p.
20).
SPR2-320.814.02.01.02
01.05
Page 21 of 46
CS SD 24