3D Reconstruction Upgrade Part 2
Fig. 35: Potentiometer
•
Move the C-arm into the -95° degree orbital position (limit switch = side of image inten-
sifier).
•
Loosen the Allen screw at the belt pulley
•
Attach the outside toothed belt (see
•
Use a screwdriver to turn the potentiometer counterclockwise into the end position
(Fig. 35 / p.
•
Once the potentiometer is in this position, turn it fully clockwise twice.
•
Secure the belt pulley with the Allen screw
Siemens AG
Medical Solutions
37).
SPR2-320.814.02.01.02
01.05
(Fig. 34 / p.
36).
(Fig. 34 / p.
36).
(Fig. 34 / p.
36).
Page 37 of 46
CS SD 24
37
ARCADIS Orbic