Agilent Technologies 5517B Operation And Service Manual page 94

Laser head
Hide thumbs Also See for 5517B:
Table of Contents

Advertisement

position. The Digital Section is OK. If A3U2C (Pin 8) doesn't go high, check divider and state machine circuitry. See
A3 Power Amp section for additional information.
s. Signature Analysis:
When taking signatures, ensure that heater test jumper is in the off position.
Damage to the high voltage power supply or laser tube assembly could result if
these instructions are not adhered to.
With the power (±15 Vdc)
follows:
Connect the Signature Analyzer (HP 5005A/B) test pod leads as follows:
Connect signature analyzer front panel controls as follows:
CAUTION
disconnected,
HTR OK = HI
-REF ON = HI
RESET = TEST
CLOCK FREQUENCY = Remove
HEATER = OFF
A3TP2, A3TP3 and Pin "X" of HTR OK = jumpered to Ground
GND = A3TP4 (labeled "GRD")
CLOCK = A3TP5 (labeled "CLK")
STOP = A3TP6 (labeled "STP")
START = A3TP7
CLOCK, START, and STOP Polarity set for falling edge.
set
the
HP5517B/C/D
switches
and
test
jumpers
as

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

5517c5517d

Table of Contents